dosimetry of an x ray tube
play

Dosimetry of an x-ray tube and Irradiation results of a DEPFET - PowerPoint PPT Presentation

Dosimetry of an x-ray tube and Irradiation results of a DEPFET matrix Carried out at the University of Karlsruhe IMPRS Workshop Motivation: The KEKB accelerator and High luminosity collider Asymetric energy for studying


  1. Dosimetry of an x-ray tube and Irradiation results of a DEPFET matrix Carried out at the University of Karlsruhe IMPRS Workshop

  2. Motivation: The KEKB accelerator and… • High luminosity collider • Asymetric energy for studying       ( 4 ) e e Y S B B • Measure CKM matrix elements • CP violation Measuring device: IMPRS Workshop 7/21/2009 2

  3. … the Belle II detector • The two innermost layers of the vertex detector will be consisted of DEPFET matrices • Ionizing Radiation from: • Beam-Gas Interaction • Synchrotron radiation will damage those matrices Ca. 1 Mrad/a (=10 kGy/a) • • Simulating of these various sources with x-ray radiation IMPRS Workshop 7/21/2009 3

  4. What are DEPFETs? • Depend on sideward depletion: Internal Gate • Generated charges will be stored in the Internal Gate • Modulation of Drain- Source current • Clear operation necessary • Beneath Gate metallization: Layer of silicon dioxide Sensitive to ionizing radiation IMPRS Workshop 7/21/2009 4

  5. Damage mechanism • Generataion of electron/hole pairs in the oxide • Mobility of holes is weak in comparison to electrons • Holes will be trapped near the interface and stay there for O(h) till O(a) • Also: Increase of interface state density IMPRS Workshop 7/21/2009 5

  6. Dosimetry of the x-ray facility in Karlsruhe • To quantify the damage done by ionizing radiation Applied Dose is necessary • Calibration was given,  Recheck it  Problem Challenge started … IMPRS Workshop 7/21/2009 6

  7. Recipe for succsesfull Calibration • X-ray source • Detector a) Fast readout, highly sensitive, good spectral response, low noise, superb amplifier b) If not (a), then find a compromise of (a) • Energy spectrum would be good a) Measure one b) Simulate one Thanks to my colleague: Oksana Brovchenko IMPRS Workshop 7/21/2009 7

  8. 1. Determine dose rate in silicon diode • Dose rate measurement via depleted diode • Measuring of reverse-current • X-ray photons generate electron/hole-pairs in Si- Bulk – Every charge carrier pair represents an energy of 3.6 eV – With the x-ray generated current one gets the deposited power, with the mass of the diode  dose rate in Si IMPRS Workshop 7/21/2009 8

  9. 2. Making use of the spectrum • Spectrum of tungsten anode (including a 0.4 mm Be filtering, black) • Generate via absorption function of Zr a new transmitted spectrum (red) IMPRS Workshop 7/21/2009 9

  10. 3. Dose in SiO 2 • Dose measurement (diode)  power of spectrum (Si, blue) is known • Original spectrum (black)  Filter (Zr, red)  Absorption in SiO 2 (green) • Final dose is determined IMPRS Workshop 7/21/2009 10

  11. New dose rates New dose rates • Every dose rate matches to a specific set of parameters, let’s assume • U=60 kV (max. tube voltage) • I=33 mA (max. tube current) 1  • Distance is 155 mm ( ) D  2 r • Dose rate in silicon (300 µm) Gy   | 0 , 305 D 60 , 33 , 155 , 300 Si kv mA mm µm s • Dose rate in silicon dioxide (180 nm) Gy   | 0 , 239 D SiO 60 kV , 33 mA , 155 mm , 180 nm 2 s IMPRS Workshop 7/21/2009 7/21/2009 11 11

  12. Back to the DEPFET: Setup and DAQ Study with a 6x16 Minimatrix • • Important contacts on PCB  easy accessibility Drain contact needed probe • needle Several irradiation and • measurement steps Readout duration of input • characteristic of all 96 pixels ~ 6…7h • – min. 4 days of room temperature annealing DAQ via LabVIEW: Sweep of • Gate voltage, Drain current is measured IMPRS Workshop 7/21/2009 12

  13. Results of input characteristics √I DS (U GS ) IMPRS Workshop 7/21/2009 13

  14. Change of threshold voltage IMPRS Workshop 7/21/2009 14

  15. Change of gain g m • Input characteristic curve fitted with – I=aU²+bU+c – Gain g m =mU+b via dI/dU  m=2a  no numeric deviation – Gain evaluated at Drain current = 50 µA • Maybe effect is part of setup and readout process  needs to be rechecked to find out if effect still occurs IMPRS Workshop 7/21/2009 15

  16. Outlook & Conclusions • Dosimetry – A good agreement between simulated spectrum and dose measurement with silicon diode has been found – Further investigations are under way: Spectrum has been measured (matches fine to the simulated one, except low energy photons • and L-Lines of tungsten) • Dosimetry with RadFETs have been done • DEPFET – Shift of threshold voltage matches to previous single pixel measurements (strongly dependent on gate voltage) – Spreading of threshold voltage after irradiation increased threshold voltage shift ∆ = ±0,12 V unirradiated • threshold voltage shift ∆ = ± 0,20 V at 142 krad (1.42 kGy) • threshold voltage shift ∆ = ±0,22 V at 2.37 Mrad (23.65 kGy) • – Impact on Belle II • Steering chips (Switcher) won‘t have problems with the spreading • Change of g m interferes on g q (via gate oxide capacitance). Variations in Drain-Source current may affect the current readout chip. IMPRS Workshop 7/21/2009 16

  17. Backup Slides IMPRS Workshop 7/21/2009 17

  18. Change of threshold voltage vs. Dose IMPRS Workshop 7/21/2009 18

Recommend


More recommend